Fluke 2680A User Manual page 181

Data acquisition system/data logging system
Hide thumbs Also See for 2680A:
Table of Contents

Advertisement

Resistance Range*
30 kΩ
30 kΩ
300 kΩ
3 MΩ
* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance plus 0.01% decade resistance tolerance. You must add any additional lead wire
resistance present in your setup to the resistance values given in this table.
Resistance Range*
30 kΩ
30 kΩ
300 kΩ
3 MΩ
* The resistance accuracy in this table makes allowance for up to 0.1
You must add any additional lead wire resistance present in your setup to the resistance values
given in this table. Using the 5700A in 2-wire compensation mode eliminates much of the lead
resistance errors. See the 5700A Operator's Manual for details
Decade Resistor
Short Circuit (Zero)
29 kΩ
290 kΩ
2.9 MΩ
5700A
Short Circuit (Zero)
19 kΩ
190 kΩ
1.9 MΩ
Accuracy Performance Tests
Minimum Reading
700 Ω
29.681 kΩ
289.07 kΩ
2.8607 MΩ
Minimum Reading
700 Ω
19.686 kΩ
189.60 kΩ
1.8740 MΩ
of lead wire resistance.
Maintenance
Maximum Reading
1 kΩ
30.019 kΩ
292.63 kΩ
2.9410 MΩ
Maximum Reading
1 kΩ
20.014 kΩ
192.10 kΩ
1.9277 MΩ
4
4-23