4. Verify Accuracy Configure the Decade Resistance Source for the output
values below and verify the Spy window measurement is between the
minimum and maximum values. Change the channel 1 range as required (see
Step 2).
Resistance Range
300 Ω
300 Ω
3 kΩ
3 kΩ
30 kΩ
300 kΩ
3 MΩ
The resistance accuracy in this table makes allowance for up to 0.01% decade resistance
tolerance.
Resistance Range
300 Ω
300 Ω
3 kΩ
3 kΩ
30 kΩ
300 kΩ
3 MΩ
Decade Resistor
Short Circuit (Zero)
290 Ω
Short Circuit (Zero)
2.9 kΩ
29 kΩ
290 kΩ
2.9 MΩ
5700A
Short Circuit (Zero)
190 Ω
Short Circuit (Zero)
1.9 kΩ
19 kΩ
190 kΩ
1.9 MΩ
Accuracy Performance Tests
Minimum Reading
0 Ω
289.861 Ω
0 Ω
2.89849 kΩ
28.9834 kΩ
289.621 kΩ
2.89146 MΩ
Minimum Reading
0 Ω
189.912 Ω
0 Ω
1.89912 kΩ
18.9893 kΩ
189.750 kΩ
1.89425 MΩ
Maintenance
Maximum Reading
0.05 Ω
290.139 Ω
0.5 Ω
2.90137 kΩ
29.0166 kΩ
290.379 kΩ
2.90854 MΩ
Maximum Reading
0.030 Ω
190.088 Ω
0.30 Ω
1.90088 kΩ
19.0107 kΩ
190.250 kΩ
1.90575 MΩ
4
4-27