Monitor Software - Nonin LS1 Operator's Manual

Lifesense ii
Table of Contents

Advertisement

Data Output and Software
PulseOx_Status, Capno_Status, and Sys_Status Bits:
PulseOx_Status
Bit
Description
23
Reserved
22
Sensor Disconnect Bit
21
Artifact Bit
20
Out of Track Bit
19
Sensor Alarm Bit
Perfusion Bits
18
'10' – red perfusion
'01" – green perfusion
17
'11' – yellow perfusion
16
Reserved
15
Reserved
14
Reserved
13
Reserved
12
Reserved
11
No SpO
2
10
Reserved
9
Reserved
8
Reserved
7
PR High Alarm Active
6
PR Low Alarm Active
5
SpO
High Alarm Active
2
4
SpO
Low Alarm Active
2
3
Reserved
2
Reserved
1
Reserved
0
Reserved
UAL is the current upper alarm limit value for each parameter.
LAL is the current lower alarm limit value for each parameter.
Alarm volume is Off, Medium or Maximum.
CRC is a 16-bit CCITT for the entire row.

Monitor Software

Visit nonin.com for more information about the monitor's software, including the latest monitor
software revision.
Capno_Status
Bit
Description
23
Reserved
22
Reserved
21
Reserved
20
Reserved
19
Reserved
18
Reserved
17
No Breath
16
No CO
2
15
Reserved
14
Reserved
13
Reserved
12
Reserved
11
Occlusion – Pump Off
10
Reserved
9
No Breath Alarm
8
Occlusion
7
EtCO
High Alarm Active
2
6
EtCO
Low Alarm Active
2
5
RR High Alarm Active
4
RR Low Alarm Active
3
Reserved
2
Reserved
1
Reserved
0
Reserved
Sys_Status
Bit
Description
7
Alarm Silence
6
Low Batt
5
Reserved
4
Reserved
3
Reserved
2
Reserved
1
Reserved
0
Reserved