Operating the Avant 9600
Option switch 7 allows users to select the duration output of a signal. In the DOWN position, a
one-second signal is output and repeated every 60 seconds when an audible alarm starts and
will continue every 60 seconds until the alarm is silenced. In the UP position, the nurse call signal
is active during the entire duration of an audible alarm.
NOTE: The Nurse Call feature overrides silenced alarms.
The Nurse Call Circuit Switch, located near the Option switch, allows users to select NO
(Normally Open—the left position) or NC (Normally Closed—the right position). The hospital's
nurse call system configuration will determine whether the Nurse Call Circuit Switch on the 9600
must be placed in the Normally Open (NO) or Normally Closed (NC).
Electrical State of Nurse Call Output
Nurse Call Switch Position
Nurse Call Not Active
Nurse Call Active
Standby Mode
Right
Open
Closed
Closed
Left
Closed
Open
Open
Nurse Call
Connection
RS232 Serial
Connection
Power
Connection
Nurse Call
Switch
1
2
3
4
5
6
7
8
Figure 3: Monitor Rear View
NOTE: It is the user's responsibility to implement the interface between the Nurse Call
system and the Avant 9600, and to adequately test the interface between the Avant 9600 and
the Nurse Call system to ensure operation.