Responsible Organization Settings Screen

This screen allows the operator to establish institution default high and low limits for pulse rate, SpO
EtCO
, and respiration rate.
2
This screen is accessed through the Configuration Menu (see Access the Configuration Menu).
NOTES:
The operator is required to enter a new 4-digit PIN the first time the Responsible Organization
Settings screen is accessed or after restoring factory defaults.
The Responsible Organization Settings are populated with factory default settings until
changed by the organization.
Once a PIN is saved, the alarm limits become restricted to the high and low limits on the
Responsible Organization Settings screen.
1. On the Configuration Menu, press Responsible Organization Settings.
2. Enter the responsible organization PIN and press OK. Responsible Organization Settings
screen displays (figure 10).
3. From this screen, the operator may:
• Set the minimum alarm volume
• Set the default alarm volume
• Set the responsible organization's alarm limits: pulse rate (PR), SpO
respiration rate (RR)
• Reset the device to the factory defaults
• Change the 4-digit PIN
4. Press Save when all changes to the Responsible Organization Settings are complete. To exit
without saving changes, press Cancel.

Figure 10. Responsible Organization Settings Screen