Battery Replacement; Spo 2 Sensors; Sample Line; Intended Use - Nonin LS1 Operator's Manual

Lifesense ii
Table of Contents

Advertisement

Battery Replacement

The battery, made of Lithium Ion (Li-Ion) rechargeable cells, is integral to the device and cannot be
replaced by anyone other Nonin Technical Service. The life expectancy of the battery is approximately
1 year.
CAUTION: Do not attempt to replace the battery inside the monitor. The battery is
!
not field replaceable and cannot be replaced by the operator. Use only Nonin-
specified components. Use of another battery may present a risk of fire or explosion.
Contact Nonin Technical Service when the battery needs replacing. Battery
replacement by inadequately trained personnel could result in a hazardous situation.
For optimal performance, the battery should be replaced once per year to limit the amount of Li build
up if the battery is charged in a cold environment.
SpO
Sensors
2
See the Accessories section for a complete list of compatible sensors. Detailed information regarding
specific sensor use (e.g., patient population, body/tissue, application) can be found in the respective
sensor instructions for use.

Sample Line

Intended Use

The sample line is used to measure the content of carbon dioxide in exhaled air (EtCO
use, disposable tubing that connects to the monitor's moisture trap with a Luer lock connector. One
sample line is included in the standard kit. The monitor can be fitted with several types of sample lines
to best suit the patient (see Accessories).
The following instructions refer to the nasal cannula supplied in the standard system configuration.
Other sample lines have separate instructions included in their packaging.

Applying the Sample Line

1. Insert the cannula into the patient's nostrils.
2. Place the tubing behind each ear.
3. Connect the Luer lock fitting to the moisture trap and twist to tighten.
WARNING: To avoid patient injury, only use Nonin-specified power supplies, cables,
and accessories (see Accessories).
CAUTION: The sample line, moisture trap, filter, and Nafion tubing are single-patient
!
use, disposable components. Do not reuse disposable accessories. Accessories
marked "single-use" must be used on one patient only and be disposed of after
usage. Dispose of all components in accordance with your local, state or national
regulations regarding waste management.
CAUTION: The patient's nasal passage may dry out if continuous monitoring is
!
required. Check patient hourly for nasal comfort.
12
System Components
). It is single-
2

Advertisement

Table of Contents
loading

Table of Contents