Alarms
Alarm
PR High Limit – displays when pulse
rate is above the high alarm limit
PR Low Limit – displays when pulse
rate is below the low alarm limit
SpO
High Limit – displays when
2
SpO
is above the high alarm limit
2
SpO
Low Limit – displays when
2
SpO
is below the low alarm limit
2
EtCO
High Limit – displays when
2
EtCO
is above the high alarm limit
2
EtCO
Low Limit – displays when
2
EtCO
is below the low alarm limit
2
RR High Limit – displays when
respiration rate is above the high
alarm limit
RR Low Limit – displays when
respiration rate is below the low
alarm limit
No breath is detected for
approximately 30 seconds
Critical Low Battery
Table 7. High Priority Alarms
Visual Indicator
PR high limit flashes red 2 times
per second.
PR low limit flashes red 2 times
per second.
SpO
high limit flashes red 2 times
2
per second.
SpO
low limit flashes red 2 times
2
per second.
EtCO
high limit flashes red
2
2 times per second.
EtCO
low limit flashes red 2 times
2
per second.
RR high limit flashes red 2 times
per second.
RR low limit flashes red 2 times
per second.
"NO BREATH" displays with a red
background that flashes 2 times
per second.
Battery indicator flashes red
2 times per second.
Audible Indicator
3 beeps, pause,
2 beeps, pause,
3 beeps, pause,
2 beeps, and a 6-second
pause.
This cycle repeats until
silenced or the alarm
condition is cleared.
35