Alarm Lock Mode Screen

This screen allows the operator to set and lock the monitor's alarm limits.
NOTES:
The locked alarm limits cannot be set to limits outside of the Responsible Organization limits.
When alarm limits are locked, limits cannot be adjusted when monitoring.
This screen is accessed through the Configuration Menu (see Access the Configuration Menu).
1. On the Configuration Menu, press Alarm Lock Mode. Alarm Lock Mode screen (figure 12)
displays.
2. To lock the alarm limits, press On. A check mark displays to the left of the selection.
3. Use each parameter's up/down arrows to adjust the high and low alarm limits.
4. When finished, press Save to save the changes, close the screen, and return to the
Configuration Menu. To exit without saving changes, press Cancel.

Calibration Screen

This screen is accessed through the Configuration Menu (see Access the Configuration Menu).

Calibration Procedure

The monitor has a built-in zero-point calibration function for CO
at least every 6 months, or if the baseline of the CO
The calibration apparatus (see Accessories) is reusable for approximately 100 times. When the
pellets start to turn purple they cannot absorb any more CO
30

Figure 12. Alarm Lock Mode Screen

2
. Perform the calibration procedure
2
graph is elevated.
and the calibration apparatus must be
2
Using the LifeSense II Monitor