Trend Screen - Nonin LS1 Operator's Manual

Lifesense ii
Table of Contents

Advertisement

Display Screens
Table 4. Operating Screen – Display Descriptions (Continued)
No.
Symbol/Name
13
Respiration Capnogram Displays a graph of the respiration and EtCO
14
15
Medium
Maximum (default)
Off

Trend Screen

The trend screen displays up to 4 hours of trending data for pulse rate, SpO
rate. The scale of the graphs is automatically set and cannot be adjusted. The 4-hour timescale is
divided into 30-minute segments.
To access the trend screen (figure 5), press the Trend icon on the operating screen.
NOTE: If a new alarm activates while the trend screen is displayed, the trend screen closes and
the display returns to the operating screen so the alarm condition is visible.
The trend screen automatically closes after 1 minute. To manually exit the trend screen, press Close.
All trend data clears when the device is turned off.
Brightness
Pressing this icon allows the operator to change the
brightness of the display while monitoring. Options are
maximum (default), medium, and low.
Audible Alarm Volume
Pressing this icon allows the operator to change the audible
alarm volume while monitoring. The minimum alarm value is
set on the Responsible Organization Settings screen.
To set the volume to Off, the operator must press OK in the
"Audible Alarms Off" pop-up window. The audible alarm may
only be shut off if allowed by the Responsible Organization
Settings.
NOTES:
Configure the alarm volume options using the
Responsible Organization Settings screen (see the
Minimum Alarm Volume and Default Alarm Volume
sections).
Audible alarms cannot be shut off if the Minimum Alarm
Volume field is set to Medium or Maximum.
The default alarm volume is either the saved Default
Alarm Volume or the last used volume (if Last Used
Alarm Setting is selected at start-up).
Description
.
2
, EtCO
2
, and respiration
2
19