Rechargeable Battery - Nonin LS1 Operator's Manual

Lifesense ii
Table of Contents

Advertisement

System Components
No. Symbol/Name
9
10
11
NONIN SpO
2
12

Rechargeable Battery

The monitor is designed to operate continuously when connected to a power outlet or for
approximately 5 hours on a fully-charged battery.
A flashing yellow battery indicator (low battery) displays when the battery is almost depleted.
This gives the operator approximately 60 minutes of use, or time to plug in the monitor before
it switches itself off.
A flashing red battery indicator (critical battery) displays when 10 minutes of power is left
before the monitor switches itself off.
Recharging a fully-depleted battery takes approximately 9 hours when using a Nonin-specified power
supply. Always connect the monitor to an outlet whenever it is not in use.
For more information, see the Internal Power section of the System Specifications.
CAUTION: Do not charge Li-Ion batteries at a temperature of 0 °C (32 °F) or less as
!
this may result in significantly reduced battery life.
WARNING: To avoid patient injury, only use Nonin-specified power supplies, cables,
and accessories (see Accessories).
Table 2. Monitor Features (Continued)
Speaker
WARNING: Ensure that all alarm volumes are audible in all
situations. Do not cover or obstruct any speaker openings.
USB Port
Connects a USB flash drive, PSG DAC cable, or Capno RTC cable to
the monitor.
CAUTION: All parts and accessories connected to the USB port of this
device must be certified according to at least IEC Standard EN 60950,
IEC 62368-1, or UL 1950 for data processing equipment.
Sensor Input
Connects the SpO
sensor to the monitor. See the Accessories section
2
for a complete list of compatible sensors. No other sensors may be
used.
WARNING: To avoid patient injury, use only Nonin-branded
PureLight pulse oximeter sensors. These sensors are
manufactured to meet the accuracy specifications of Nonin pulse
oximeters. Using other manufacturers' sensors can result in
improper pulse oximeter performance.
Power Supply Input
Connects the power supply to the monitor. Only use Nonin-specified
power supplies.