Reset Device to Factory Defaults
This setting discards all user settings and returns the monitor to the factory default settings and alarm
limits.
Restore Factory Defaults
1. Access the Responsible Organization Settings screen.
2. Press Reset Device to Factory Defaults.
3. In the "Factory Restore" pop-up window:
• Press Yes to restore the factory defaults. The monitor:
- Reverts to the factory default minimum alarm volume, default alarm volume, and alarm
limits (see table 6).
- Deletes the PIN.
- Removes the Responsible Organization Settings limit restrictions.
• Press No to cancel.
New PIN
NOTE: The operator is required to enter a new 4-digit PIN the first time the Responsible
Organization Settings screen is accessed or after restoring factory defaults.
The PIN, which may be set to any 4-digit number, is used to unlock parameter settings on the
Responsible Organization Settings screen.
Change the PIN
1. Access the Responsible Organization Settings screen.
2. Press New PIN.
3. In the "Enter New PIN" pop-up window, enter a 4-
digit PIN.
NOTE: If the PIN is less than four digits, the "Invalid
PIN Entered" pop-up window displays. Press OK to
acknowledge and press New PIN again.
4. Press OK to save the new PIN. To exit without
saving changes, press Cancel.
5. Display returns to the Responsible Organization
Settings screen.
Factory Service
This feature is only used by Nonin Medical, Inc.