Deep Sleep Mode

If the monitor is not plugged in and has been in Standby mode for more than an hour, it enters Deep
Sleep mode.
The monitor immediately enters Deep Sleep mode if the battery level at shut down is either low (yellow
battery indicator) or critical (red battery indicator).

System Setup

1. Place the monitor in a position so the display can be clearly seen.
• If using the monitor portably or in an environment without power, the monitor can operate for
approximately 5 hours on a fully-charged battery. The battery indicator on the display shows
the battery capacity.
• If using the monitor while connected to the power supply and a power outlet, the green
charging indicator is lit. The battery indicator on the display shows a lightning bolt to indicate
the battery is charging.
2. Visually inspect the monitor and make sure it has no visible signs of damage.
3. Examine the sensor for obvious defects. Ensure the sensor is clean if it has been previously
used.
4. Connect the sensor to the SpO
5. Replace the single-use, disposable moisture trap and filter before each use. Refer to Single-
Patient Use, Disposable Moisture Trap and Filter for instructions on how to replace the
moisture trap and filter.
CAUTION: After exposing the monitor to an environment outside of normal room
!
temperature/humidity conditions, always replace the moisture trap and filter before
each use.
6. Connect the sample line to the connector on the moisture trap and secure it by turning the
Luer lock connector clockwise. Only use sample lines recommended by Nonin (see
Accessories).
7. See Monitoring a Patient.
WARNING: Prior to connecting the monitor to the power supply and a power outlet, be
sure to verify the voltage and frequency rating on the power supply are the same as
the outlet. If this is not the case, do not connect the monitor and power supply to the
outlet.
WARNING: The monitor displays a flashing yellow battery indicator (low battery) when
it has approximately 60 minutes of use remaining before it shuts itself off.
22
port located on the side of the monitor.