Maintenance and Inspection

Maintenance and Inspection

Maintenance

Ensuring Optimal Performance

In order to ensure safety and optimal performance of the monitor, Nonin recommends a yearly
inspection and functional check be performed on the monitor (see Recommended Inspections and
Functional Check). The inspection and functional check may be performed by Nonin Technical
Services or at your facility.
Perform the calibration procedure at least every 6 months, or if the baseline of the CO
elevated (see Calibration Procedure). After calibration, the monitor should be verified using 5% CO
gas. The calibration apparatus, gas valve, and 5% CO
Accessories).
Please contact Nonin Technical Service if monitor maintenance cannot be performed at your facility.
The Oxitest
Plus7
(software rev. 2.5 or greater) by Datrend Systems, Inc. can be used to verify operation
of the pulse oximeter.

Cleaning the Sensor

Refer to individual sensor Instructions for Use for details.

Cleaning the Monitor

1. Clean the monitor with a soft cloth moistened with isopropyl alcohol. Do not use any cleaning
solution other than what is recommended here, as permanent damage could result.
2. Dry with a soft cloth, or allow to air dry.
CAUTION: Always turn off the monitor prior to cleaning the monitor. Do not place the
!
device in liquid or clean it with agents containing ammonium chloride or bleach. Do
not sterilize or autoclave the monitor or accessories.
verification gas are available from Nonin (see
2
graph is
2
2
43