Testing Summary; Spo 2 Accuracy Testing; Pulse Rate Accuracy Testing; Low Perfusion Accuracy Testing - Nonin LS1 Operator's Manual

Lifesense ii
Table of Contents

Advertisement

Specific to this example, the response of the 4 beat-average is 1.5 seconds.
CAUTION: A functional tester cannot be used to assess the accuracy of a pulse
!
oximeter monitor or sensor.

Testing Summary

SpO
accuracy and low perfusion testing were conducted by Nonin Medical, Inc., as described below:
2
SpO
Accuracy Testing
2
SpO
accuracy testing is conducted during induced hypoxia studies on healthy, non-smoking, light-
2
to-dark-skinned subjects during motion and no-motion conditions in an independent research
laboratory. The measured arterial hemoglobin saturation value (SpO
arterial hemoglobin oxygen (SaO
oximeter. The accuracy of the sensors in comparison to the co-oximeter samples measured over the
SpO
range of 70 – 100%. Accuracy data is calculated using the root-mean-squared (A
2
all subjects, per ISO 80601-2-61, Standard Specification for Pulse Oximeters for Accuracy.

Pulse Rate Accuracy Testing

This test measured pulse rate oximeter accuracy with and without motion artifact simulation
introduced by a pulse oximeter tester. This test determines whether the oximeter meets the criteria of
ISO 80601-2-61 for pulse rate during simulated movement, tremor, and spike motions.

Low Perfusion Accuracy Testing

This test uses an SpO
2
at various SpO
levels for the oximeter to read. The oximeter must maintain accuracy in accordance
2
with ISO 80601-2-61 for pulse rate and SpO
(0.3% modulation).
56
) value, determined from blood samples with a laboratory co-
2
Simulator to provide a simulated pulse rate, with adjustable amplitude settings
2
at the lowest obtainable pulse amplitude
Technical Information
) of the sensors is compared to
2
rms
value) for