Cautions - Nonin LS1 Operator's Manual

Lifesense ii
Table of Contents

Advertisement

Indications for Use
Warnings (Continued)
The monitor displays a flashing yellow battery indicator (low battery) when it has approximately 60 minutes of use remaining
before it shuts itself off.
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device
should be observed carefully to verify normal operation.
Prior to connecting the monitor to the power supply and a power outlet, be sure to verify the voltage and frequency rating on
the power supply are the same as the outlet. If this is not the case, do not connect the monitor and power supply to the outlet.
The use of accessories other than those specified in this manual may result in increased electromagnetic emission and/or
decreased immunity of this device.
As with all medical equipment, carefully route cables and cannula to reduce the possibility of entanglement, strangulation, or
tripping.
Refer to the applicable sensor instructions for use for additional warnings and cautions.

Cautions

!
LifeSense II should only be operated by trained licensed practitioners.
To prevent damage to the monitor, operate and store the monitor in an upright position.
Visually inspect the monitor for damage before each use. Do not use a damaged monitor or system.
Verify display functionality before each use.
Setting alarm limits to extremes can render the alarm system useless.
Each time the system is turned on, audible alarms are silenced for 2 minutes unless the operator presses the Audio Pause
button.
When an alarm is acknowledged and audible alarms are paused, new alarms will have visual indicators, but not an audible
indicator.
Capnography alarms are not active until the first breath is detected.
Oximetry alarms are not active until the first pulse is detected.
To ensure the longevity and safety of the monitor, the carrying case must be used when operating the monitor in an emergency
medical services (EMS) environment.
Do not mount the monitor directly above the patient. If the monitor is mounted, be sure to check that the adjustable mounting
clamp is securely affixed.
When mounting the monitor to a mobile pole, mounting the monitor higher than 1.5 meters (5 feet) or mounting more than
2 kilograms (4.5 pounds) of equipment onto the pole may result in tipping, damage to the equipment, or injury.
Do not use cables longer than 3 meters (10 feet).
Always turn off the monitor prior to cleaning the monitor. Do not place the device in liquid or clean it with agents containing
ammonium chloride or bleach. Do not sterilize or autoclave the monitor or accessories.
Do not simultaneously touch the accessible connector pins and the patient.
After exposing the monitor to an environment outside of normal room temperature/humidity conditions, always replace the
moisture trap and filter before each use.
The sample line, moisture trap, filter, and Nafion tubing are single-patient use, disposable components. Do not reuse
disposable accessories. Accessories marked "single-use" must be used on one patient only and be disposed of after usage.
Dispose of all components in accordance with your local, state or national regulations regarding waste management.
Set or adjust alarm parameters one at a time.
The oximeter sensor might not work on cold extremities due to reduced circulation. Warm or rub the finger to increase
circulation, or reposition the sensor.
2

Advertisement

Table of Contents
loading

Table of Contents