Alarms
Inoperable Alarms
Inoperable alarms (table 9) are medium priority alarms. To correct the condition, perform these steps:
1. Turn the monitor off and then back on again to remove the error message.
2. If the error persists, note the error code and contact Nonin Technical Service at (800) 356-
8874 (USA and Canada), +1 (763) 553-9968, or +31 (0)13 - 79 99 040 (Europe).
Message/Error Code
NO OXIMETER E03
NO CAPNO E04
SPEAKER FAULT E05
SYSTEM FAILURE E06
BUTTON FAILURE E07
MEMORY CORRUPT E08
Low Priority Alarms
Low priority alarms do not have an audible alarm.
Alarm
Low perfusion.
A poor pulse signal was detected.
Table 9. Inoperable Alarms
No communication from the pulse oximetry unit.
No communication from capnography unit.
Speaker is not working properly.
General system error.
Audio Pause button is damaged.
Error detected in memory structure.
Table 10. Low Priority Alarms
"LOW PERFUSION" displays with a cyan
background.
The cyan
Probable Cause
Visual Indicator
indicator displays.
37