Default Settings

The Avant 9600 features Factory Default and User-Defined Default settings.
WARNING: The battery pack must be installed at all times while the device
is operating—even when operating on AC power. If it is
necessary to operate the device without batteries, audible
alarms and memory functions may not be available. DO NOT
use the device without batteries when patient safety relies
upon audible alarms.
Factory Default Setting
In Factory Default setting, all adjustable alarm and volume parameters are set at
their default values. Factory Default setting is the Avant 9600's default operating
setting. It is indicated by option switch 4 in the DOWN position.
User-Defined Default Setting
In User-Defined Default Setting (option switch 4 in the UP position), alarm limit
and volume settings must be adjusted. When this setting is first activated, valid limit
settings must be entered for SpO
volume, and pulse rate volume; the Avant 9600 will not return to normal operating
mode until all default values have been set. Once set, the adjusted values are used as
defaults until the Avant 9600 is turned on with option switch 4 in the DOWN
position, at which time the device restarts in Factory Default setting.
NOTE: Patient Security mode overrides default settings.
NOTE: In User-Defined Default setting, the lower SpO
cannot be set below 80.
alarm limits, pulse rate alarm limits, alarm
2
alarm limit
2
12