Pulse Rate Alarm Limit Settings

1. Upon entering the pulse rate low alarm limit, "H L" will appear in the SpO
rate low alarm limit may be set from 30 to 110 (beats per minute) in increments of 5, or to "0FF".
2. After selecting the pulse rate low alarm limit, the SpO
setup mode for the pulse rate high alarm limit. The pulse rate high alarm limit may be set from 75
to 275 (beats per minute) in increments of 5, or to "0FF".
3. When the desired pulse rate high alarm limit setting has been selected, setup mode continues to
the SpO
high alarm limit.
2
SpO
High Alarm Limit Settings
2
1. Upon entering the SpO
alarm limit may be set from 80 to 100 (%) in increments of 1, or to "0FF".
2. When the SpO
high alarm limit setting has been selected, setup mode continues to the breath
2
beep pitch settings.

Settings

Breath Beep Pitch
Upon entering the breath beep pitch, "FPt" will appear in the SpO
defaults to "variable pitch" (fixed pitch off). Refer to "Audible Breath Beep" on page 26 for additional
information. The breath beep pitch selection may be toggled between "ON" or "OFF".
Calendar Settings
NOTE: Setting the month to "00" disables the calendar and clock functions. The calendar and
clock functions are used to time stamp real-time data for memory. Unless you intend to use real-
time data output or memory playback options, skip this section.
1. After the calendar setting has been selected in the setup mode, "y" will appear in the SpO
display indicating the calendar setup mode for the year. The year may be set to "00" through
"99".
2. After selecting the year, the display will show "nn" indicating the setup mode for the month. The
month may be set to "00" through "12".
3. After selecting the month, the display will show "d" indicating the setup mode for the day of the
month. The day may be set to "01" through "31". After selecting the day, the setup mode
continues to the clock settings.
Clock Setting
1. After the calendar settings have been selected in the setup mode, "h" will appear in the SpO
display indicating clock setup mode for the hour. The hour may be set to "00" through "23".
2. After selecting the hour, the SpO
minutes. The minutes may be set to "00" through "59".
3. After selecting the minutes, the device will:
a. exit the setup mode,
b. briefly display "rCL ALr" (recall alarm settings) and then the current alarm limit settings, and
c. begin normal operation.
20
high alarm limit, "02H" will appear in the SpO
2
display will show "nn" indicating the setup mode for the
2
Using the 9847 Pulse Oximeter
2
display will show "H H" indicating the
2
display. The SpO
2
display. The breath beep pitch
2
display. The pulse
high
2
2
2