Table 1. Start-Up Screen Icons And Display Descriptions - Nonin RespSense LS1R-9R Operator's Manual

Capnography monitor
Table of Contents

Advertisement

Table 1: Start-up Screen Icons and Display Descriptions
No.
Name
Audible Alarm Disable Pressing this icon turns the audible alarms off.
1
Audible Alarm Enable Pressing this icon turns the audible alarms on.
2
30kg/66lbs
3
>30kg/66lbs
4
RespSense Version
5
Software Revision
6
Level
SW: REV
It disables the audible alarms by setting all lower limits to 0.
Default if no icon is chosen.
Pressing this icon selects the default alarm limits for patients
weighing 30 kg (66 lbs) or less.
Only available if Audible Alarm Enable icon is chosen on
previous screen.
Pressing this icon selects the default alarm limits for patients
weighing more than 30 kg (66 lbs).
Only available if Audible Alarm Enable icon is chosen on
previous screen.
Shows RespSense version.
If an error occurs during start-up, an error number displays here
and an alarm activates.
Shows the software revision level installed on the RespSense
monitor.