Repair and Service
Machine specific tools, software tools and auxiliary means
2.6.1.1
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.2/38
Agfa Company Confidential
Note 2: How to save the " DICOM, LPD and HIPAA report"
In the "Service Engineer Tools" go to the "DICOM, LPD and HIPAA report"
page – right mouse key – save page as.
How to view the logfiles offline
Step #
Descpription
Step 1
Install the PIDEX
tool
Step 2
Extract the
corresponding log
file of the session
file.
Step 3
In the PIDEX tool
select "Tools –
logview" and open
the desired logfile
Step 4
In the PIDEX tool go
to function "Edit –
find" to find the
desired information.
A session file is named e.g.
DS5302_1052_SES_21OCT2005_09-55-48.ZIP, where the different figures
represent following information:
DS5302:
Printer (Drystar 5302)
1052:
Serial Number
SES:
Session File
21OCT2005: Current date
09-55-48:
Current time HH-MM-SS
Interesting for troubleshooting are following files within the SES files:
devil.log (controller and print engine);
time.log
Printed copies of this document are not controlled.
(Type 5366/100)
Comment
Available on MEDNET, GSO Library
See note below
It is also possible to select a single
word in the logfile. Then all other
occurances of this expression are also
highlighted.