AGFA DRYSTAR 5302 Technical Documentation Manual page 217

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
HC311003 : inputsensormodule.s4
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC311004 : inputsensormodule.s13
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC311005 : inputsensormodule.sensors
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W311003
None
S4 not connected
Connect S4
S4 bracket module defect
Replace
MOD11
Failure
W311004
None
S13 not connected
Connect S13
S4 bracket module defect
Replace
MOD11
Failure
W311005
None
one of the inputsensors is not functioning
Remove jam