AGFA DRYSTAR 5302 Technical Documentation Manual page 214

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
HC309020 : powersupplymodule.powerth
Type
Context
Action
Reason
Cure
Part
HC309022 : powersupplymodule.powermotor
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC309024 : powersupplymodule.powerlogic
Type
Context
Action
Reason
Cure
Part
HC309025 : powersupplymodule.power3v3
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
09
Replace
PSU module defect
Replace
MOD09
Defect
09
Replace
Actuators power cables short circuited
Check actuators power cables
PSU module defect
Replace
MOD09
Defect
09
Replace
PSU module defect
Replace
MOD09
Defect
09
Replace
PSU module defect
Replace
MOD09
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 58 of 72

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd+dis150.05e

Table of Contents