DD+DIS150.05E
HC308020 : devilmodule.power.sensor
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC308021 : devilmodule.p42.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC312000 : tyndall.temperature
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E308020
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
Defect
E308021
SeeManual
Connector P42 not connected
Connect P42
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
Defect
E312000
SeeManual
Ambient temperature too high what could be caused by continuous printing
in an already warm environment or the environment temperature is just too
high
If caused by continuous printing , this can be cured by waiting until
temperature comes down ; otherwise it has to be made sure that the
environment temperature comes down
Cooling of MOD12 CPU board (Themis or Tyndall) not efficient because of
blocked airflow
Make sure that enough airflow is available for the MOD12 CPU board
(Themis or Tyndall) by insuring that air inlets and outlets are completely
free
Defect CPU on MOD12
DOCUMENT CONTROL NOTE: