DD+DIS150.05E
2.3

Print Head (TH) Calibration

The Print Head Profile calibration is necessary whenever vertical stripes are not longer
When to be
performed
acceptable:
Vertical stripes are unacceptable in case they are visible in the medical service test
image.
REQUIRED TOOLS:
One of following, calibrated densitometers: McBeth TR-924, Gretag D200-2, X-
Rite 301, X-Rite 310, X-Rite 331, X-Rite 341
Lint free cloth
Isopropyl - alcohol
(1)
Print the Medical service test image.
Instructions
(2)
Clean the thermal head with a cleaning tissue lightly moistened with Isopropyl -
alcohol. The last cleaning action must be carried out in one way.
(3)
In the webpages - menu 'Service Engineer Tools - Enter Service Command
session' select '6. Calibration - 3. Print Head Calibration'.
(4)
Follow the instructions on the WebPages. Refer to Figure 4.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Chapter 3.6 / 11
Agfa Company Confidential