AGFA DRYSTAR 5302 Technical Documentation Manual page 215

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
HC309026 : powersupplymodule.power12v
Type
Context
Action
Reason
Cure
Part
HC309027 : powersupplymodule.powern12v
Type
Context
Action
Reason
Cure
Part
MODULE : displaykeypadmodule
Elabel : Yes
HC310000 : displaykeypadmodule.elabel.present
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
09
Replace
PSU module defect
Replace
MOD09
Defect
09
Replace
PSU module defect
Replace
MOD09
Failure
W310000
None
Display Elabel not connected
Connect Display elabel
Display Elabel defect
Replace
MOD10