Adjustments

Following adjustments have to be performed at the printer:
Adjustment
Horizontal film
alignment
1.1

Horizontal Film Alignment Procedure

The horizontal film alignment is necessary when:
When to be
performed
The TPH has been replaced and
the blank frame at the left and right hand side on a SMPTE test image is < 1,5
mm for DT2B / DT2C Films.
Centering the image on the film: Compensate for mechanical tolerances of the
Purpose
thermal head mounting.
REQUIRED TOOLS:
Ruler to measure in the range up to 15 mm
REQUIRED TIME:
10 min.
NOTE:
The horizontal film alignment can be performed via browser user interface or the local
keypad. The required time and proceeding is identical.
It can be performed at any film size.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
To be done:
In case after TH replacement the blank left /
right frame on a SMPTE test image is < 1,5 mm
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Adjustments and Calibrations
Agfa Company Confidential
Details see:
1.1
Chapter 3.6 / 4