Advertisement

DD+DIS150.05E
LIST OF ERRORS
MODULE : uppertraymodule
Elabel : No
HC301000 : uppertraymodule.connected
Type
Context
Action
Reason
Cure1
HC301001 : uppertray.overruled
Type
Context
Action
Reason
Cure
HC301002 : uppertray.disabled
Type
Context
Action
Reason
Cure
HC301003 : uppertraymodule.tagreader
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E301000
SeeManual
XS7 or XS8 not connected
Connect XS7 or XS8
Failure
W301001
None
Inputtray overruled (dmax 2.0)
Replace filmpack
Failure
W301002
None
Inputtray disabled
Replace filmpack
Failure
W301003
None
Tag reader not connected
Connect tag reader
Tag reader defect
Replace
Tag reader