Known Problems; Software Release 1.8.0_C2; Solved Software Bug - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
1.3

Known Problems

Software 1.8.0_C1 has following known software problems:
In case the Online Diagnosis detects that the Output Module is Causeive and has
to be exchanged, it does not display "MOD 06 - Please replace module" but "E06 -
Please replace module".
The Module pages in the Service Engineer Tools are not yet operational
In case a film jam happens with last film in a package, the printer does not show
"empty film tray" but continues printing with the "protective cover", what leads to
jam with the protective cover.
Automatic film calibration fails in case of a pending maintenance action.
2

Software Release 1.8.0_C2

Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
 A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
 On MedNet always the latest available software release can be downloaded.
2.1

Solved Software Bug

Thermal Head calibration could lead to vertical stripes on the film in the range of
Symptom
 0.05 O.D*. units (For comparison: Under best viewing conditions the human eye can
distinguish 0.01 O.D. units).
*O.D. = Optical Density
Software fault
Cause
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Software Releases and Patches
Software Update
11/2005
 SN1156 plus 1143 and SN 1149
MedNet, GSO Library
None
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Chapter 3.8 / 6
Agfa Company Confidential