DD+DIS150.05E
2
2.1
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Machine specific safety instructions and hints for repair work
Safety check upon completion of the repair work
Upon completion of any repair job, especially after replacing or installing
conductive parts, check the following points as a routine safety check.
Country specific regulations must be observed as well.
Safety check
Visual check
Electrical check
Visual check
• Make sure that the power cord and the cables inside the machine do not
show any signs of damage.
• Make sure that the ventilation slots before the print head and the power
supply are free and that the print head and power supply ventilation is
working properly.
• Make sure, that no obstacles can cause stumbling (e.g. cables), and no
defective covers, sharp edges or parts sticking out in the vicinity of the
machine can cause injuries.
• Make sure that no smoldering spots or wrong cabling is visible.
• Make sure that all grounding connections to metallic covers are present.
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
Repair and Service
Details see..
Chapter 3.1 / 5
Agfa Company Confidential
2.1
2.2