Guidelines Regarding Safety During Service Work; General Safety Guidelines For Service Work; Safety Guidelines For Printed Circuit Board Replacements - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
1.3
1.3.1
1.3.2
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Machine specific safety instructions and hints for repair work

Guidelines regarding safety during service work

This section describes the safety guidelines to be considered during service
work.
Safety precaution

General safety guidelines for service work

Safety Guidelines for Printed Circuit Board Replacements

Safety Guidelines regarding Service PC Connection
Safety Guidelines regarding the Thermal Print Head
General safety guidelines for service work
The following points must be observed during all repair or maintenance work:
Make sure to always disconnect the machine to be repaired or maintained
from the mains.
Caution: Even when the machine is switched off, as long as it has not
been disconnected from the mains some components may still be
supplied.
Caution: If the mains connection is absolutely necessary for the
respective action (repair and maintenance work), this job must only be
done by trained personnel.
Before printer service or replacing conductive parts always switch off the
printer and disconnect the power.
Before switching the printer off make sure that there are no films being
printed at this time.
Replace defective parts with original spare parts only.
Safety Guidelines for Printed Circuit Board Replacements
When handling printed circuit boards (abbr.: PCBs) the following points must
be observed:
• Always switch off the printer and unplug the power cord, before you
disconnect or connect cables on printed circuit boards.
• When working on PCBs, always wear an anti-static wrist strap. Never
touch any parts or components on PCBs with your bare fingers.
• PCBs have to be kept or transported in their protection bags. Never carry
a PCB without protection bag and walk on carpet or plastic floor covering
(electrostatic charge).
Once the PCB is taken out of its protection bag, it has to be protected
from electrostatic charge by a grounded mat.
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
Repair and Service
Details see..
1.3.1
1.3.2
1.3.3
1.3.4
Chapter 3.1 / 3
Agfa Company Confidential

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd+dis150.05e