Print Head Profile Calibration - AGFA DRY 3.4 Reference Manual

Hide thumbs Also See for DRY 3.4:
Table of Contents

Advertisement

Print head profile calibration

To start the manual print head profile calibration:
1. Press the Down key once to select 'Printhead profile', and press the
Confirm key.
The 'Select input tray' screen appears:
2. Press the Up/Down arrow keys to select an input tray, followed by the
Confirm key (e.g. choose upper input tray).
The system will start with Phase 1 of the calibration: Registration:
As a result, the 'Registration' print out appears.
3. After registration, the system will continue with Phase 2 of the calibration:
the print head profile calibration:
As a result, the 'TPH profile' print out appears.
4. In case the print head profile calibration has failed, the following message
appears:
DRY 3.4 | Advanced Operation (key-operator mode) | 63
WARNING:
Print head profile calibration must be done whenever the profile
of the print head needs to be improved (manual print head
profile calibration).
Note: In the procedure, we will assume a print head profile
calibration on the upper input tray. The procedure for the lower
input tray is identical.
2604A EN 20180322 1454

Advertisement

Table of Contents
loading

This manual is also suitable for:

5364/500

Table of Contents