Page 1
HealthCare Maintenance Order-No.: DD+DIS015.02E *1UT4861* ADC Compact Plus Type 5146 / 100/200 1 Piece UT486 MA1 Edition ADC Compact Plus Digitizer Type 5146 / 100/200 Base: cPCI This documentation is separately available. Order No: DD+DIS015.02E printed in Germany 01 / 2002...
Page 2
The CE Declaration (CE Conformity) becomes invalid if the product is changed without explicit consent of the manufacturer! This applies to all parts, not only to safety elements. We reserve the right to technical changes Chapter 9 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
Checking the Image Quality........12 Test Cycles................12 Exposure of a Flatfield ............12 Evaluation of a Flatfield ............15 Completion of Maintenance ........16 Maintenance Checklist Edition 1, Revision 1 ADC Compact Plus Chapter 9 / I (Type 5146 / 100/200)
As is the case for all technical devices, the ADC Compact Plus must be operated, cared for and serviced correctly. · If you don't operate the ADC Compact Plus correctly or if you don't have it serviced correctly, Agfa-Gevaert is not liable for resulting disturbances, damages or injuries.
Insert the cassette in the input buffer as described in the User Manual. At all times, keep your fingers clear of the input slot. As soon as the ADC Compact Plus takes in the cassette, release it.
2.2 Throughput For throughput most important are the cycles per day. They usually count between 50 and 200. Chapter 9 / 4 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
Troubleshoot these errors with the help of the technical documentation, chapter 3.3, “Troubleshooting”. 5.9 Error List Total 3.1.4 Clear Infocounter (1) Clear the infocounters to refresh relative counters. Edition 1, Revision 1 ADC Compact Plus Chapter 9 / 5 (Type 5146 / 100/200)
In any case replace the roller once a year. Figure 2 Belt (3) Exchange the transport belt once a year. Chapter 9 / 6 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
(2) Open the doors when the machine is switched on. (3) The power supply of the cPCI- Rack and its fan must still be Figure 5 Edition 1, Revision 1 ADC Compact Plus Chapter 9 / 7 (Type 5146 / 100/200)
If it is very dirty, remove it and clean the brushes 1 in the direction of the bristle - do not bend. Figure 7 Chapter 9 / 8 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
(CM+9.0450.6553.0) (1) Open the doors of the digitizer. (2) Clean the grid of the fan with a soft cloth and a dust brush. Figure 9 Edition 1, Revision 1 ADC Compact Plus Chapter 9 / 9 (Type 5146 / 100/200)
Page 13
Do not touch the lamps with bare fingers! Use a soft cloth to insert the lamp. Figure 10 (9) Re-insert the Erasure Unit. Chapter 9 / 10 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
Visual check (1) Check if there are scratches on the surface (2) Check if edges are loose as an indication for mechanical problems at IP transport Edition 1, Revision 1 ADC Compact Plus Chapter 9 / 11 (Type 5146 / 100/200)
<Send flatfield> <Calibration pattern> <Banding pattern> Print the flat fields "Calibration" and "Banding" via the Processing Station (window setting of 0.6, without changing the level setting). Chapter 9 / 12 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
Page 16
In the <Sub-examination> list, click <Flat field>. · Confirm the Exposure class <200>. Make sure that the outlined areas are filled in as shown in the example. Edition 1, Revision 1 ADC Compact Plus Chapter 9 / 13 (Type 5146 / 100/200)
Page 17
(4) Insert the cassette into the digitizer and print the image on a printer with a window setting of 0.6 without changing the level setting. Figure 12 Chapter 9 / 14 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
(see beside). · Check polygon monitoring entries and diagnostic images. If there are still unacceptable effects please contact the Support Center. Figure 14 Edition 1, Revision 1 ADC Compact Plus Chapter 9 / 15 (Type 5146 / 100/200)
(2) Make a backup of the system on floppy (3) Inform the customer about what was done during the maintenance and which repairs need to be done in next future. Chapter 9 / 16 ADC Compact Plus Edition 1, Revision 1 (Type 5146 / 100/200)
Page 20
AGFA and the Agfa-Rhombus are trademarks of Agfa-Gevaert AG, Germany Herausgegeben von / Published by / Edité par / Pubblicato da / ../ Editado por Agfa-Gevaert AG Fototechnik Tegernseer Landstraße 161 D - 81539 München Printed in Germany / Imprimé en Allemagne /...
Page 21
HealthCare Maintenance Checklist Order-No.: DD+DIS015.02E ADC Compact Plus Type 5146 / 100/200 Work Instructions for order no. Cycles The maintenance has to be carried out: · Once a year or every 25 000 cycles Maintenance must be carried out according to the maintenance instructions, see chapter 9 of your Service Documentation.
General · Confirm the maintenance · Make a backup of the system on floppy · Inform the customer about maintenance and which repairs are necessary Maintenence Checklist ADC Compact Plus Edition 1, Revision 1 Page 2 (Type 5146 / 100/200)
Page 23
DD+DIS015.02E Maintenance Checklist Remarks: ........................................................................................................................................................................................................................................... Date / Signature Service Technician Customer Edition 1, Revision 1 ADC Compact Plus Maintenence Checklist (Type 5146 / 100/200) Page 3...