C
3 D
HAPTER
ISASSEMBLY
Removing the Defibrillator Boards and Capacitor Bank
Modules
Tools Required
#1 Phillips screwdriver with torque capability
•
#2 Phillips screwdriver with torque capability
•
26 mm open end wrench
•
Large diagonal cutters
•
5k Ohm, 25 watt, resistor
•
Setup
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
•
Complete these procedures in the following order before performing this procedure:
•
1. "Removing the Handle" on page 61
2.
3.
To reinstall the defibrillator boards and/or capacitor bank modules, reverse
the steps.
Defibrillator Boards and Capacitor Bank Disassembly
1
Lay back end of separated
unit down with opened side
facing up.
2
Pull the black snap rivet
straight out.
86
P
ROCEDURES
"Removing the Printer Module" on page 62
"Disassembling the Main Unit" on page 64
9650-0807-01 Rev. C