C
3 D
HAPTER
ISASSEMBLY
Removing the Dock Connector
Tools Required
#1 Phillips screwdriver with torque capability
•
Setup
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
•
Complete these procedures in the following order before performing this procedure:
•
1. "Removing the Handle" on page 61
2.
3.
4.
Dock Connector
1
Remove 4 screws and
washers from the back panel
that are securing the dock
connector to the housing.
Note: During reassembly,
torque screws to 6 in-lbs.
2
Disconnect black connector
from the dock connector
board.
94
P
ROCEDURES
"Removing the Printer Module" on page 62
"Disassembling the Main Unit" on page 64
"Removing the Defibrillator Boards and Capacitor Bank Modules" on page 86
9650-0807-01 Rev. C