9.0 Advisory Test

Fluke Impulse 7000 DP Analyzer (or equivalent), ECG simulator
Equipment
Do This...
9.1
Connect the multifunction cable (MFC) to
the Analyzer.
9.2
Turn on the Propaq unit and connect the
pads.
9.3
On the Analyzer, select VF (ventricular
fibrillation), then press the Analyze button.
9.4
On the Propaq unit, press the CHARGE
button to activate, then press the CHARGE
button again.
9.5
On the Propaq unit, press the Shock button. Verify that the shock count is
9.6
On the Analyzer, select NSR (normal sinus
rhythm), then press the Analyze button.
9650-0807-01 Rev. C
ZOLL Propaq MD Service Manual
Maintenance Tests
Observe This...
The unit displays Pads on the top
left of the screen.
The Propaq unit displays:
ANALYZING
SHOCK ADVISED
The unit charges to the selected
energy level.
displayed at the top of the screen.
The Propaq unit displays:
ANALYZING
NO SHOCK ADVISED
Pass/Fail
o o
o o
o o
o o
o o