Shock Test - ZOLL Propaq MD Service Manual

Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE

8.0 Shock Test

WARNING!
Take the necessary precautions to guard against shock or injury before you start
conducting the defibrillator tests.
Keep hands and all other objects clear of the multi-function cable connections and
defibrillator analyzer when discharging the defibrillator.
Before you discharge the defibrillator, warn everyone near the equipment to STAND
CLEAR.
Fluke Impulse 7000 DP Analyzer (or equivalent), stop watch
Equipment
Do This...
8.1
Turn on the Analyzer.
8.2
On the Analyzer, select the DEFIB function.
8.3
On the Analyzer, select ENERGY.
8.4
Connect the Propaq MFC connector to the
pads connection.
8.5
On the Propaq unit, apply power to the unit
using a battery or an AC adapter.
8.6
Connect the MFC cable to the Propaq unit.
8.7
Turn on the Propaq unit, and press the Leads
quick access key
8.8
On the Propaq unit, set energy level to 5J by
pressing the Energy Select DOWN arrow.
8.9
On the Propaq unit, press the CHARGE
button.
8.10 On the Propaq unit, press the Shock button.
8.11 On the Analyzer, observe and record the
energy measurement.
8.12 On the Propaq unit, set the energy level to
50J by pressing the Energy Select UP arrow.
8.13 On the Propaq unit, press the CHARGE
button.
22
T
ESTS
I, II,
(
)
to select
PADS
III...
Observe This...
.
The unit charges to 5J, the red
Shock button illuminates and a
ready tone sounds.
Energy is delivered at 5J. Verify
that the delivered energy is within
15% based on the impedance
values of your analyzer in the
table "Delivered Energy at 200J
Defibrillator Setting into a Range
of Loads" on page 24.
The unit charges to 50J, the red
Shock button illuminates and a
ready tone sounds.
Pass/Fail
o o
o o
o o
9650-0807-01 Rev. C