C
1 M
HAPTER
AINTENANCE
25.0 Temperature Test
IBP Temperature Simulator (or equivalent)
Equipment
Do this...
25.1 Connect the temperature simulator to the
Propaq T1 channel.
25.2 On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
25.3 Repeat steps 1 and 2 of this procedure for the
T2 channel.
25.4 Disconnect cable.
44
T
ESTS
Observe This...
The displayed temperature value is
within 1.0 F or 0.5 C of the
simulator setting.
9650-0807-01 Rev. C
Pass/Fail
o o