Etco2 Calibration Check - ZOLL Propaq MD Service Manual

Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE
17.0 EtCO
Calibration Check
2
5% CO
Equipment
2
Products, PN 4653ORF-BD)
Note:
Do This...
17.1
On the
Supervisor>Service>Diagnostic>CO
Calibration to enter the CO
control panel.
17.2
Connect the
calibration gas cylinder using the sample
line included with the calibration kit.
17.3
From the CO
select Verify Calibration.
17.4
Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.
17.5
Continue to apply the calibration gas until
the status in the CO
panel displays: CALIBRATION GAS CAN
BE REMOVED.
17.6
Verify that the unit displays:
VERIFICATION DONE - OK.
17.7
Verify that the measured gas reading is
between 4.8 and 5.2 (Vol%).
17.8
Exit out of calibration by pressing the Back
arrow key (
Display/Home button (
34
T
ESTS
calibration gas cylinder with CO
If you have previously run the CO2 Calibration procedure, start at step 3.
unit, select
Propaq
Propaq
unit to the 5% CO
Calibration control panel,
2
Calibration control
2
) three times or pressing the
).
sample line (kit available from Scott Medical
2
Observe This...
>
2
Calibration
2
2
Pass/Fail
o o
o o
o o
9650-0807-01 Rev. C