Removing the Defibrillator Board and Capacitor Bank
Modules
Tools Required
#1 Phillips screwdriver with torque capability
#2 Phillips screwdriver with torque capability
26 mm open end wrench
Setup
This procedure should be conducted on a static mat.
Wear a wrist strap to ground during the procedure.
Complete these procedures in the following order before performing this procedure:
1.
2.
3.
Note:
To reinstall the defibrillator board and/or capacitor bank modules, reverse
the steps.
Defibrillator Board and Capacitor Bank Disassembly
1
2
3
4
9650-0801-01 Rev B

Removing the Defibrillator Board and Capacitor Bank Modules

"Removing the Handle" on page 57
"Removing the Printer Module" on page 58
"Disassembling the Main Unit" on page 60
After replacing the Defibrillator Board and/or Capacitor Bank modules, see Chapter 4,
"Resetting Life Cycle Counters" to reset the life cycle counter for the module.
Lay back end of separated unit down with
opened side facing up.
Remove interface board and disconnect
the black cable by pulling the black snap
rivet straight out.
Remove the multi-color cable from the
defibrillator board.
Pull out small interconnect board.
ZOLL Propaq MD Service Manual
multi-color cable
73