C
1 M
HAPTER
AINTENANCE
This chapter describes the following maintenance tests:
1.0 Physical Inspection of the Unit
•
2.0 Operational TestsOperational Tests
•
3.0 Leads Test
•
4.0 Heart Rate Display Test
•
5.0 Heart Rate Alarm Test
•
6.0 Power Supply Test
•
7.0 Leakage Current Test
•
8.0 Shock Test
•
9.0 Advisory Test
•
10.0 Synchronized Cardioversion Test
•
11.0 Pacer Test
•
12.0 Printer Test
•
13.0 Treatment Report Test
•
14.0 SpO2 Test (for SpO
•
15.0 SpO2, SpCO, and SpMet Test (for SpO
•
16.0 EtCO2 Calibration (required for annual calibration)
•
17.0 EtCO2 Calibration Check
•
18.0 EtCO2 Test
•
19.0 IBP Test
•
20.0 NIBP Calibration Check
•
21.0 NIBP Leak Test
•
22.0 Deflation Test
•
23.0 Inflation Test
•
24.0 NIBP Test
•
25.0 Temperature Test
•
26.0 Wi-Fi Test (Optional)
•
Before You Begin the Maintenance Tests
Assemble the tools listed in following section, "Equipment You Need to Perform the
•
Keep an extra fully charged SurePower II battery available.
•
Schedule an hour to conduct the entire maintenance test.
•
Perform the tests in the order presented.
•
Perform all the steps of each test procedure.
•
Complete all the steps of the procedure before evaluating the test results.
•
6
T
ESTS
feature without SpCO and SpMet)
2
Maintenance Tests."
feature with SpCO, or SpCO and SpMet)
2
9650-0807-01 Rev. C