20.0 NIBP Calibration Check

NIBP Simulator, dual lumen hose w/ NIBP test fixture cuff (9100-3072-TF) or equivalent
Equipment
Note:
Setup
Attached the hose connected to the T-fitting of the NIBP test fixture cuff to the pressure port of
the NIBP simulator. Set simulator to Static Pressure.
Do This...
20.1 Turn on the Propaq unit.
Note: If the unit is in AED mode, press
the Manual quick access key to go into
manual mode.
20.2 Connect a dual lumen hose from the NIBP
simulator to the NIBP port of the Propaq.
20.3 On the Propaq unit, do the following to
enter the NIBP Test control panel:
Note: You need a supervisor passcode to
enter the Supervisor menu.
20.4 From the Propaq NIBP Test control panel,
use the navigation keys to select:
20.5 Monitor the pressure reported by the
external pressure standard.
20.6 Once the pressure reaches the target
pressure of 50 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.7 Allow the system to stabilize for at least
10 seconds.
20.8 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
20.9 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
9650-0807-01 Rev. C
If ECG simulator is connected to the device, either disconnect ECG simulator or
disable the Smart Cuff feature in the Supervisor Menu.
Press the More (
) then the Setup
(
) quick access keys.
Use the navigation keys to select
Supervisor>Service>Device
Tests>NIBP Test.
Close Valve
Turn Pump On
ZOLL Propaq MD Service Manual
Maintenance Tests
Observe This...
Reading should be within
3mmHg.
Reading should be within
4mmHg.
Pass/Fail
o o