Etco2 Calibration (Required For Annual Calibration) - ZOLL Propaq MD Service Manual

Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

16.0 EtCO
Calibration (required for annual calibration)
2
5% CO
Equipment
2
Products, PN 4653ORF-BD)
Do This...
16.1
Turn on the
16.2
On the
Supervisor>Service>Diagnostic>CO
Calibration to enter the CO
control panel.
16.3
Connect the
calibration gas cylinder using the sample
line included with the calibration kit.
16.4
From the CO
Calibration.
16.5
Activate the calibration gas flow by
pressing the top valve on the calibration gas
canister.
16.6 Continue to apply the calibration gas until the
status in the CO
displays: CALIBRATION GAS CAN BE
REMOVED.
The calibration date will be updated upon
exiting the CO
9650-0807-01 Rev. C
calibration gas cylinder with CO
unit.
Propaq
Propaq
unit, select
unit to the 5% CO
Propaq
Calibration Menu, select
2
Calibration control panel
2
When the calibration is complete, the
status in the CO
Calibration control panel
2
displays: CALIBRATION DONE-OK
Test control panel.
2
ZOLL Propaq MD Service Manual
sample line (kit available from Scott Medical
2
Observe This...
>
2
Calibration
2
2
Maintenance Tests
Pass/Fail
o o
o o
33