10.0 Synchronized Cardioversion Test

Fluke Impulse 7000DP Analyzer (or equivalent)
Equipment
Connect the cable to the defibrillator analyzer.
Test Setup
Select Sync mode on analyzer.
Do this...
10.1 Verify that the Propaq unit is turned on
and the
10.2 Press the SYNC softkey on the
defibrillator. Enter the synchronized
cardioversion timing test mode on the
defibrillator analyzer.
10.3 Select 200J.
10.4 Press the CHARGE button. When the
Shock button illuminates, press and
hold it.
26
T
ESTS
Input 1mV ECG signal at 60 BPM.
lead is selected.
PADS
Observe this...
Sync appears on display.
Sync markers display on the monitor.
The sync marker appears as an S over
the ECG R-wave peaks on strip chart
and display.
Observe that the R-wave to shock delay
(sync delay) is less than 60 milliseconds
on the analyzer display.
Defibrillator discharges.
Pass/Fail
o o
o o
9650-0807-01 Rev. C