7.0 Leakage Current Test
See the manufacturer's instructions or supplied specifications for the leakage tester you use.
Equipment
See the manufacturer's instructions or supplied specifications for the leakage tester you use.
Setup
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.
ECG
MFC
Earth Leakage
9650-0807-01 Rev. C
Maximum Leakage Acceptance Limits
Patient Leakage
Normal
Single Fault
10
50
10
50
Normal
2500
ZOLL Propaq MD Service Manual
Mains on Applied Part
Single Fault
100
100
Single Fault
5000
Maintenance Tests
21