15.0 SpO
, SpCO, and SpMet Test
2
(for SpO
feature with SpCO, or SpCO and SpMet)
2
Note:
Fluke Index 2 SpO
Equipment
(Masimo P/N 3379) or Masimo Rainbow 3 ft Direct Connect Optical Test Sensor
Disconnect any ECG cable or IBP probes from the
Test Setup
Do This...
15.1 Power on the Propaq unit without the
SpO
15.2 Connect the
unit.
15.3 Set the simulator to 75%
BPM. Attach the
simulator.
15.4 Set the simulator to 75%
BPM.
9650-0807-01 Rev. C
Use this test if your unit has SpO
have the SpMet feature on the unit, disregard the SpMet values in the test below.
Simulator (or equivalent), Masimo Rainbow 3 ft BioTek Test Sensor
2
sensor attached.
2
SpO
sensor to the Propaq
2
SpO
SpO
sensor to the
2
SpO
ZOLL Propaq MD Service Manual
and SpCO, or SpO
2
Propaq
unit.
Observe This...
Verify that the
displays INITIALIZING, then
goes blank.
Verify that
SpO
displays SEARCHING.
Verify that after a brief period of
and 50
2
initialization,
SpO
The
numeric displays an
2
SpO
value between 72.1% and
2
78%.
The
SpCO
numeric displays an
SpCO
value between 13.5 and
21.5.
The
SpMet
numeric displays an
SpMet
value between 5.5 and
9.5.
Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.
Verify that after a brief period of
and 230
2
initialization,
The
SpO
numeric displays an
2
SpO
value between 72.1% and
2
78%.
SpCO
The
numeric displays an
SpCO
value between 12.8 and
20.8.
SpMet
The
numeric displays an
SpMet
value between 5.2 and
9.2.
Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.
Maintenance Tests
, SpCO and SpMet. If you do not
2
Pass/Fail/N/A
SpO
numeric
2
numeric
2
o o
o o
o
o o
o o
o o
o
o o
o
o
o