23.0 Temperature Test
IBP Temperature Simulator (or equivalent)
Equipment
Do this...
23.1 Connect the temperature simulator to the
Propaq MD T1 channel.
23.2 On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
23.3 Connect the Temperature simulator to the
Propaq MD T2 channel
23.4 On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
23.5 Disconnect cable.
9650-0801-01 Rev. B
Propag MD Service Manual
Propaq MD Overview
Observe This...
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
Pass/Fail
o o
o o
43