Spo2 Test - ZOLL Propaq MD Service Manual

Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE
13.0 SpO
Fluke Index 2 SpO
Equipment
Disconnect any ECG cable or IBP probes from the Propaq MD unit.
Test Setup
Do This...
13.1 Power on the Propaq MD unit without the
SpO
13.2 Connect the
unit.
13.3 Set the simulator to 70%
Attach the
13.4 Set the simulator to 80%
13.5 Set the simulator to 90%
13.6 Disconnect the
30
T
ESTS
Test
2
Simulator (or equivalent), Nellcor™ SPO2 Finger Probe
2
sensor attached.
2
SpO
sensor to the Propaq MD
2
SpO
2
SpO
sensor to the simulator.
2
SpO
2
SpO
2
SpO
simulator.
2
Observe This...
Verify that the
displays INITIALIZING, then
goes blank.
Verify that
displays SEARCHING.
Verify that after a brief period of
and 60 PPM.
initialization, the
displays an
67% and 73%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
and 60 PPM. Verify that after a brief period of
initialization, the
displays an
77% and 83%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
and 60 PPM. Verify that after a brief period of
initialization, the
displays an
87% and 93%.
Verify that the displayed pulse
rate reads 60 +/- 3 bpm.
Pass/Fail
SpO
numeric
2
SpO
numeric
2
o o
SpO
numeric
2
SpO
value between
2
o o
o o
SpO
numeric
2
SpO
value between
2
o o
o o
SpO
numeric
2
SpO
value between
2
o o
9650-0801-01 Rev. B