Synchronized Cardioversion Test - ZOLL Propaq MD Service Manual

Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE

9.0 Synchronized Cardioversion Test

Fluke Impulse 7000DP Analyzer (or equivalent)
Equipment
Connect the cable to the defibrillator analyzer.
Test Setup
Select Sync mode on analyzer.
Do this...
9.1 Verify that the Propaq MD unit is
turned on and the
selected.
9.2 Press the SYNC softkey on the
defibrillator. Enter the synchronized
cardioversion timing test mode on the
defibrillator analyzer.
9.3 Select 200J.
9.4 Press the CHARGE button. When the
Shock button illuminates, press and
hold it.
26
T
ESTS
Input 1mV ECG signal at 60 BPM.
lead is
PADS
Observe this...
Sync appears on display.
Sync markers display on the monitor.
The sync marker appears as an S over
the ECG R-wave peaks on strip chart
and display.
Observe that the R-wave to shock delay
(sync delay) is less than 60 milliseconds
on the analyzer display.
Defibrillator discharges.
Pass/Fail
o o
o o
9650-0801-01 Rev. B

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents