3.0 Leads Test

3-lead, 5-lead, and 12-lead cables, Fluke Impulse 7000DP Analyzer
Equipment
Note:
Connect the lead wires appropriate for each test to the
Test Setup
equivalent, then connect the ECG cable to the Propaq MD unit.
Do this...
3.1 Turn on the Propaq MD unit.
Note: Make sure that the unit is not
in PADS.
3.2 Disconnect one lead from the
simulator.
3.3 Reconnect the lead. Repeat step 2
with the remaining leads.
3.4 Repeat steps 2 and 3 for 5 lead and 12
lead cables.
9650-0801-01 Rev. B
Test each cable separately.
Propag MD Service Manual
Fluke Impulse 7000DP Analyzer
Observe this...
No ECG LEAD FAULT message
displayed.
The ECG LEAD FAULT message displays
within 3 seconds (if configured).
Wait for ECG LEAD FAULT message to
clear from the display (if configured).
Note: If heart rate alarm sounds, on the
Propaq MD unit:
Press the More quick access key (
to access the second set of quick access
keys.
Press the Alarm quick access key (
Press the Alarm Cancel quick access key
(
).
Propaq MD Overview
or
Pass/Fail
o o
o o
o o
o o
) key
).