Reading Out Data; Evaluating Data - JUMO LOGOSCREEN Operating Manual

Paperless recorder with compactflash card as storage medium
Hide thumbs Also See for LOGOSCREEN:
Table of Contents

Advertisement

2 Instrument description
Optimization of
recording
duration

2.9 Reading out data

Memory alarm
Read-out via
interface

2.10 Evaluating data

26
The recording duration can be optimized by process-oriented selection of the
storage cycle.
In normal operation (no error, no alarm, ...) a storage cycle that is as long as
possible (e.g. 60sec, 180sec, ...) should be selected, depending on the
particular application.
In the event of an alarm or error, the storage cycle can be shortened via event
operation, with the effect that the measurement data are recorded with a high
time resolution.
In addition to automatic read-out via the external CompactFlash memory card,
measurement data can also be read out through the serial interface or the
Ethernet interface.
Both read-out options (card/interface) work in parallel. For this reason, there
are also two "digital signals", which indicate when the available storage space
has fallen below a certain configurable value.
The following limits can be configured at the configuration level:
Configuration
Device data
Configuration
Device data
Configuration
Device data
The parameter Parameterization
which storage space display is shown in the status line.
For further information, please refer to:
v Chapter 2.8 "Storing data"
Chapter 2.4 "Digital signals"
Chapter 3.3 "Parameterization" (page 57)
Chapter 4 "Configuration parameters" (page 73 and page 76)
Please use the PC software PCC to read out measurement data via the serial
or the Ethernet interface. The software has been developed especially for the
paperless recorder.
v Please refer to the Operating Manual B 70.9702.0. for further information.
Please use the PC Evaluation Software (PCA3000) for evaluating the data on
the PC. The software has been developed especially for the paperless
recorder.
v Please refer to the Operating Manual B 70.9701.0. for further information.
Memory alarm
Int.mem./ser.read-out
Memory alarm
Int.mem./read-out via CF
Memory alarm
CF card (external)
Memory display can be used to determine

Hide quick links: